DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
73 of 180
12101 - SLOW SCAN a velocity malfunction during scan
12102 - SLOW SCAN stopped during scan
12103 - SLOW SCAN stopped during scan
12104 - SLOW SCAN: Timeout while moving to SENSOR S9
Description: The SLOW SCAN AY exceeded the speed limit when moving the SCAN
PLATE.
Cause 1: The speed of the SLOW SCAN AY exceeded the limit. A band might appear on
the image.
Action: Do the Slow Scan Current and Velocity Tests.
Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: 1. Check for an obstruction.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: 1. Check for an obstruction.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The SLOW SCAN AY did not reach the EXTRACTION BAR SENSOR
S9.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals