DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
69 of 180
11021 - DUPLEX CAM did not move to “LATCH” position
11022 - DUPLEX CAM did not move to “UNPINCH” position
SLOW SCAN AY
12001 - SLOW SCAN EXTRACTION BAR did not move to Home Position
Cause 1: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause 2: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause 1: The DUPLEX CAM malfunctioned.
Action: Do the checkout for the CASSETTE HANDLING AY.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause 1: The EXTRACTION BAR malfunctioned.
Action: 1. Remove any obstructions in the path of the EXTRACTION BAR.
2. Check the SLOW SCAN MOTOR.
3. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
4. Check the EXTRACTION BAR HOME POSITION
Cause 2: The EXTRACTION BAR SENSOR S9 malfunctioned.
Action: Do the checkout for the EXTRACTION BAR SENSOR S9.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals