DIAGNOSTICS Troubleshooting
10DEC05
DG4825-1
Page
119 of 180
Section 3: Troubleshooting
Initialization Errors
“Power-On Self-Test” (POST) Errors
Important
An error occurs if the CR 825/850 SYSTEM energizes but does not display the main
menu.
In normal operation, the computer in the CR 825/850 SYSTEM emits one beep after
initializing.
The computer emits more than one beep when a malfunction occurs.
If the TOUCH SCREEN displays a POST Error, see the “Troubleshooting Section” of the
Owner’s Manual for the IBM COMPUTER.
1 Listen to the beeps when the computer energizes.
2 Use the Table 9 to diagnose the malfunction.
Table 9 Beep Codes
Beep Code Malfunction
1-2 Failure of the video ROM
1-1-3 Write/read test failure of the CMOS
1-2-2-3 Checksum failure of the BIOS ROM
1-2-1 Failure of the PROGRAMMABLE INTERVAL TIMER
1-2-2 Failure to initialize the DMA
1-2-3 Write/read test failure of the DMA PAGE REGISTER
1-2-4 “Refresh verification” failure of the RAM
1-3-3-1 Failure of the “First 64K RAM” test
1-3-2 Failure of the “First 64K RAM parity” test
2-2-3-1 Failure of the “Interrupt vector loading” test

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals