DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
62 of 180
10017 - Eject CASSETTE: Retry attempt
10019 - Eject CASSETTE: Timeout error
Description: The CASSETTE LOAD SENSOR S2 did not clear when the DRIVE
ROLLERS started to eject the CASSETTE.
Cause 1: The CASSETTE LOAD SENSOR S2 operates correctly, but an obstruction
occurred.
Action: Check:
DRIVE ROLLERS rotate freely
CASSETTE LOAD SENSOR S2 for obstructions
adjustment of the SLED CAM FOLLOWER
Cause 2: The CASSETTE LOAD SENSOR S2 malfunctioned.
Action: Check:
status of the CASSETTE LOAD SENSOR S2 - use Individual Component
Control
LED DS21 on the MOTION SYSTEM CONTROL BOARD A1 changes status
power to the CASSETTE LOAD SENSOR S2
condition of the EMITTER and DETECTOR of the CASSETTE LOAD SENSOR
S2
Description: The CASSETTE LOAD SENSOR S2 did not clear in the correct time
when the DRIVER ROLLERS ejected the CASSETTE
Cause 1: The CASSETTE LOAD SENSOR S2 operates correctly, but an obstruction
occurred.
Action: Check:
DRIVE ROLLERS rotate freely
CASSETTE LOAD SENSOR S2 for obstructions
adjustment of the SLED CAM FOLLOWER

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals