DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
70 of 180
12002 - EXTRACTION BAR REFERENCE SENSOR S9 failed to activate
12003 - PLATE PRESENT SENSOR S5 did not detect PLATE
Cause 1: The EXTRACTION BAR SENSOR S9 operates correctly, but another malfunction
occurred.
Action: 1. Remove any obstructions from the path of the EXTRACTION BAR.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Cause 2: The EXTRACTION BAR SENSOR S9 malfunctioned.
Action: Do the checkout for the EXTRACTION BAR SENSOR S9.
Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunction
occurred.
Action: 1. Check:
• PLATE was removed from the CASSETTE
• SLOW SCAN MOTOR operates correctly
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
3. Remove any obstructions from the path of the EXTRACTION BAR.
Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.
Action: Do the checkout PLATE PRESENT SENSOR S5.