DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
59 of 180
10008 - CASSETTE REAR SENSOR S3 did not clear
Cause 2: The CASSETTE REAR SENSOR S3 malfunctioned.
Action: Check:
status of CASSETTE REAR SENSOR S3 - use Individual Component
Control
LED DS22 on the MOTION SYSTEM CONTROL BOARD A1 changes status
power to the CASSETTE REAR SENSOR S3
condition of the CASSETTE REAR SENSOR S3
Cause 1: The CASSETTE REAR SENSOR S3 operates correctly, but another malfunction
occurred.
Action: Check:
DRIVE ROLLERS rotate freely
CASSETTE REAR SENSOR S3 for obstructions
adjustment of the SLED CAM FOLLOWER
Cause 2: The CASSETTE REAR SENSOR S3 malfunctioned.
Action: Check:
status of the CASSETTE REAR SENSOR S3 - use Individual Component
Control
LED DS22 on the MOTION SYSTEM CONTROL BOARD A1 changes status
power to the CASSETTE REAR SENSOR S3
condition of the CASSETTE REAR SENSOR S3