DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
82 of 180
12907 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12908 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12911 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12913 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
Description: The FINE NUDGE DOWN command did not occur. The command is
executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The PLATE did not move to the CASSETTE. The command is executed
again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The SOFT RESET command did not occur. The command is executed
again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The EXTRACTION BAR did not move to the mechanical adjustment
position one. The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals