DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
71 of 180
12004 - PLATE PRESENT SENSOR S5 failed to activate
12009 - Caution! SLOW SCAN MOTOR drawing too much power
Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunction
occurred.
Action: 1. Check:
PLATE was removed from the CASSETTE
SLOW SCAN MOTOR operates correctly
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
3. Remove any obstructions from the path of the EXTRACTION BAR.
Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.
Action: Do the checkout PLATE PRESENT SENSOR S5.
Cause 3: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause: A malfunction occurred in one or more of the following components:
COIL BOARD
EXTRACTION BAR SENSOR S9
SLOW SCAN AY
Action: 1. Check for an obstruction.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.