14002 - STEPPER: Malfunction of 24 V Power
14003 - CAM MOTOR M1: Low Power Malfunction
14005 - Front DOOR Open
Description: The 24 V DC CIRCUIT BREAKER of the STEPPER is open.
Cause: Malfunction of the 24 V DC power distribution.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The MOTOR-LOW CIRCUIT BREAKER is open.
Cause 1: The CAM MOTOR M1 has a short circuit.
Action: Check the wires to the CAM MOTOR M1.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause 1: The INTERLOCK SWITCH S8 malfunctioned.
Action: Do the checkout INTERLOCK SWITCH S8.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.