DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
72 of 180
12012 - SLOW SCAN AY: Timeout Error
12013 - SLOW SCAN AY: Timeout error
12014 - MSC BOARD A1: Software Error for SLOW SCAN
12100 - SLOW SCAN CONTROLLER BOARD A6: Software Error
Description: The PLATE did not move within the correct time.
Cause: A malfunction occurred in one or more of the following components:
COIL BOARD
SLOW SCAN AY
Action: 1. Check for an obstruction.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: A command for the PLATE that is not correct was made by the system
software.
Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: This error might occur during the diagnostics if the service person starts
to move the PLATE before the PLATE stops moving.
Cause: The SLOW SCAN AY malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The SLOW SCAN AY will not move.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: 1. Energize and de-energize the CR 825/850 SYSTEM.
2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals