DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
92 of 180
15013 - Image Quality Warning
15014 - LASER power not calibrated
15015 - PLATE not calibrated
Description: The image quality might be wrong. The MASTER CENTRAL PROCESSING
UNIT BOARD A2 received a system error from the MOTION SYSTEM
CONTROL BOARD A1 and was automatically repaired. If a CASSETTE was in
the CR 825/850 SYSTEM when the error occurred, the TOUCH SCREEN
displays a message about a problem with the image.
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause 3: A light leakage occurred.
Action: Do the Light Test.
Cause 4: The INTERLOCK SWITCH S8 malfunctioned. The DOOR might be open during
scan.
Action: Do the checkout INTERLOCK SWITCH S8.
Cause: The LASER DIODE malfunctioned.
Action: Do the calibration for the LASER DRIVER BOARD A17. See OPTICS MODULE.
Cause: The CR 825/850 SYSTEM does not have the calibration data for all sizes of
PLATES.
Action: Do the Calibration for the CR 825/850 SYSTEM.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals