DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
113 of 180
47520 - 47522 - Internal communications failure to UPS1
47550 - The BATTERY in UPS1 is too low. System will be shutdown in 1
minute
47551 - Unable to establish communications with UPS1
47950 - Display only - no Log Message
47951 - Display only - no Log Message
Cause: Communication error.
Action 1: 1. Check the RS-232 CABLE to COM3.
2. Do the checkout for the UNINTERRUPTED POWER SUPPLY UPS1.
Cause 1: Malfunction of the utility power.
Action: Check the POWER PLUG and wait for the power to begin.
Cause 2: The level of the BATTERY is 25% or less of the maximum level.
Action: Do the checkout for the UNINTERRUPTED POWER SUPPLY UPS1.
Cause 1: Malfunction of the utility power.
Action: Check the POWER PLUG and wait for the power.
Cause 2: The level of the BATTERY is 25% or less of the maximum level.
Action: Do the checkout for the UNINTERRUPTED POWER SUPPLY UPS1.
Description: The UNINTERRUPTED POWER SUPPLY UPS1 detects a utility power
failure condition that did not clear within 20 seconds.
Cause: Malfunction of the utility power.
Action: Check the POWER PLUG.
Description: The UNINTERRUPTED POWER SUPPLY UPS1 detects that the
BATTERY charge low is indicated from the UNINTERRUPTED POWER
SUPPLY UPS1.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals