DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
60 of 180
10009 - CASSETTE LOAD SENSOR S2 did not clear
Cause 1: The CASSETTE LOAD SENSOR S2 operates correctly, but an obstruction
occurred.
Action: Check:
DRIVE ROLLERS rotate freely
CASSETTE LOAD SENSOR S2 for obstructions
adjustment of the SLED CAM FOLLOWER
Cause 2: The CASSETTE LOAD SENSOR S2 malfunctioned.
Action: Check:
status of the CASSETTE LOAD SENSOR S2 - use Individual Component
Control
LED DS21 on the MOTION SYSTEM CONTROL BOARD A1 changes status
power to the CASSETTE LOAD SENSOR S2
condition of the EMITTER and DETECTOR of the CASSETTE LOAD SENSOR
S2