DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
64 of 180
10022 - MSC BOARD A1: Software malfunction
10024 - MSC BOARD A1: Software malfunction
10029 - READER detected the CASSETTE BAR CODE label as backward
10030 - READER did not read the CASSETTE BAR CODE
Description: The DRIVE ROLLER received a signal when waiting for an interrupt or
timeout to occur.
Cause 1: The software for the MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Install the software again for the MOTION SYSTEM CONTROL BOARD A1.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: A timeout occurred setting up the motion for the DRIVE ROLLER.
Cause 1: The software for the MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Install the software again for the MOTION SYSTEM CONTROL BOARD A1.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Cause 1: The CASSETTE was inserted backward into the CR 825/850 SYSTEM.
Action: Insert the CASSETTE correctly.
Cause 2: The BAR CODE LABEL is installed backward on the CASSETTE.
Action: 1. Remove the BAR CODE LABEL.
2. Install a new BAR CODE LABEL.
Cause 1: The CASSETTE:
has a bar code that you cannot read
does not have a bar code
Action: Install the correct bar code.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals