Page #81 background imageLoading...
Page #81 background image
DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
81 of 180
12903 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12904 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12905 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12906 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
Description: The SLOW SCAN AY did not move to the home position. The command
is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The COARSE NUDGE UP command did not occur. The command is
executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The COARSE NUDGE DOWN command did not occur. The command is
executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Description: The FINE NUDGE UP command did not occur. The command is
executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals