12941 - Unexpected Retry during Communication to MOTION SYSTEM
CONTROL BD
ERASE LAMPS
13001 - 13008 ERASE LAMP PAIRS
Description: A message could not be sent to the SLOW SCAN CONTROLLER
BOARD A6. The command is executed again.
Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL
BOARD A7.
Code Log Message
13001 ERASE LAMP PAIR No. 1 is out
13002 ERASE LAMP PAIR No. 2 is out
13003 ERASE LAMP PAIR No. 3 is out
13004 ERASE LAMP PAIR No. 4 is out
13005 ERASE LAMP PAIR No. 5 is out
13006 Two ERASE LAMP PAIRS are not functioning
13007 Three ERASE LAMP PAIRS are not functioning
13008 Four ERASE LAMP PAIRS are not functioning
Cause: One or more ERASE LAMP PAIRS malfunctioned.
Action: Install a new ERASE LAMP PAIR.