DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
89 of 180
DATA PATH
15001 - MCPU BOARD A2: Fatal Error in Software
15002 - MCPU BOARD A2: Fatal Error in Software
15003 - MCPU BOARD A2: Communications Error in Software
15005 - LASER: Reference Error
Description: A PSOS system error occurred that it is not possible to repair. The
“Event Log” includes the PSOS system error number, module, and line
number.
Cause: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Description: A system software error occurred. The “Event Log” will include the PSOS
system error number, module, and line number.
Cause: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Description: A timeout occurred for a communication between the MASTER
CENTRAL PROCESSING UNIT BOARD A2 and the MOTION SYSTEM
CONTROL BOARD A1.
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.
Description: The reference voltage of the LASER is less than 50% of the original
value.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals