DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
99 of 180
16016 - Malfunctioning PMT
16017 - Possible Light Leak
16018 - Possible Light Leak
16019 - Calibration could not find offset, rotate GALVO counterclockwise
16020 - Calibration could not find offset, rotate GALVO clockwise
Description: A PMT does not match the requirements of the diagnostic test.
Cause: The PMT/DAS BOARD A5 malfunctioned.
Action: Do the checkout for the PMT/DAS BOARD A5.
Cause: The DOOR is open, a PANEL is loose, or a light leakage occurred in the
CASSETTE HANDLING SYSTEM.
Action: 1. Do the Light Test to locate the source of the light leakage.
2. Repair the light leakage.
Description: An error occurred during “AUTO-ZERO.” The PMTs detected a large
quantity of light.
Cause: The DOOR is open, a PANEL is loose, or a light leakage occurred in the
CASSETTE HANDLING SYSTEM.
Action: 1. Do the Light Test to locate the source of the light leakage.
2. Repair the light leakage.
Cause: The calibration could not locate the “Offset.
Action: Rotate the GALVO counterclockwise.
Cause: The calibration could not locate the “Offset.
Action: Rotate the GALVO clockwise.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals