DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
63 of 180
10020 - Malfunction at CASSETTE ENTRY SENSOR S1
10021 - MSC BOARD A1: Software malfunction
Cause 2: The CASSETTE LOAD SENSOR S2 malfunctioned.
Action: Check:
status of the CASSETTE LOAD SENSOR S2 - use Individual Component
Control
LED DS21 on the MOTION SYSTEM CONTROL BOARD A1 changes status
power to the CASSETTE LOAD SENSOR S2
condition of the EMITTER and DETECTOR of the CASSETTE LOAD SENSOR
S2
Description: The CASSETTE ENTRY SENSOR S1 is blocked when it should be clear.
Cause 1: The CASSETTE ENTRY SENSOR S1 malfunctioned.
Action: Check:
status of the CASSETTE ENTRY SENSOR S1 - use Individual Component
Control
LED DS19 on the MOTION SYSTEM CONTROL BOARD A1 changes status
power to the CASSETTE ENTRY SENSOR S1
condition of the EMITTER and DETECTOR of the CASSETTE ENTRY
SENSOR S1
Description: The DRIVE ROLLER received a command that was not correct.
Cause 1: The software for the MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Install the software again for the MOTION SYSTEM CONTROL BOARD A1.
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

Table of Contents

Other manuals for Kodak DirectView CR 825

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kodak DirectView CR 825 and is the answer not in the manual?

Kodak DirectView CR 825 Specifications

General IconGeneral
BrandKodak
ModelDirectView CR 825
CategoryMedical Equipment
LanguageEnglish

Related product manuals