DIAGNOSTICS Error Codes
10DEC05
DG4825-1
Page
116 of 180
Installation
55100 - Error on the Options Diskette or this is not a valid Options
Diskette
55101 - Could not open files on the Options Diskette. Try another Options
Diskette
55102 - Pentium III identification not enabled on this Pentium III class
machine
Note
This error does not display on a Pentium 4 PROCESSOR.
Cause: A DISKETTE that is not an OPTIONS DISKETTE is installed.
Action: Install a new OPTIONS DISKETTE.
Cause: The file on the OPTIONS DISKETTE malfunctioned.
Action: Install a new OPTIONS DISKETTE.
Cause: The serial number of the Pentium III PROCESSOR is not enabled.
Action: Enable the serial number of the Pentium III PROCESSOR.