DIAGNOSTICS Using the Diagnostics
10DEC05
DG4825-1
Page
44 of 180
5 Check that the SENSORS match the table during the correct action.
6 If the SENSORS do not match the table, check the error log. See Checking the Error
Logs.
7 See the Error Codes for the possible cause and action for the problem.
Action S1 S2 S3 S5 S8 S9
A8 CAM
Inner
A8 CAM
Outer
Loading CASSETTE X X X X
Loading PLATE X X X X X
Returning PLATE to CASSETTE X X X X X
Ejecting CASSETTE X X X X X X