Introduction
Laborie NXT Pro Owner’s Manual LBL-00146 [C], MAN2010 16
Laborie equipment and accessories are licensed by governments, approved by safety
agencies, and warranted to work only with one another.
The Roam NXT and the PIM NXT are not intended for long-term direct patient contact. If
the devices must be attached to a patient, place gauze or another material between the
devices and the patient’s skin.
Caution: United States federal law restricts this device to sale or use by or on the order
of a licensed physician.
Warnings:
The NXT Urodynamic System contains magnets that can interfere with the control of a
pacemaker or implanted cardioverter-defibrillator (ICD). Follow the instructions from the
manufacturer of such devices and ensure that the Roam NXT, Urocap NXT, and charger
cable connectors are always kept at a sufficient distance from any implanted device.
Only Laborie-trained technicians may service the unit.
Do not use the NXT system if the device packaging has been opened, damaged, or if it
presents any fault due to improper transport, storage, or handling that could in any way
hamper its use. The NXT system is intended for installation by trained personnel ONLY.
Ensure that all wheels are locked during testing procedures to prevent injury to the
patient. Do not lean on the equipment or hang excessively heavy objects on the platform;
only hang Laborie standard accessories on the platform.
Use caution when adjusting the PC and monitor from its lowest or highest position;
ensure that all extremities are clear of moving parts. If the PC or monitor is removed
from the mount, the unit can rise abruptly as the gas spring triggers used are calibrated
to handle the weight of the original configuration.
Do not immerse the NXT system or any of its modules in water or any other liquids.
Do not place your fingers or any body parts inside the pump when the pump rollers are
moving.
Do not use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous
oxide.
Do not use electromyography (EMG) simultaneously with high-frequency surgical
equipment.
Do not operate the NXT system near shortwave or microwave therapy equipment, as
proximity may produce instability in the APPLIED PARTS.
Do not allow applied parts linked to a PATIENT CONNECTION, but not connected to the
PATIENT, to contact other conductive parts including those connected to protective earth.
Preserve the patient connection electrical isolation.
Do not reuse single-use accessories. After use, dispose of single-use accessories in
accordance with local regulations.
Reuse, reprocessing, or resterilization of disposables can lead to device failure and create
a risk of cross-infection and/or cross-transmission of infectious disease(s) from one
patient to another.
NXT system accessories are not intended to be defibrillation-proof.