Maintenance and Service
Laborie NXT Pro Owner’s Manual LBL-00146, MAN2010 97
System. Laborie is not responsible for any virus-related computer problems after delivery of
the computer to the customer.
5.6 Disposing of Product After Use
The NXT system equipment and disposables should be disposed of in the following manner:
Contaminated single-use disposables: Discard these products according to
the standard operating procedures on medical waste handling at the institution of
use.
Waste Electrical and Electronic Equipment (WEEE): Ensure that any waste
electrical and electronic equipment is collected separately and returned to the
locally designated recycling service for these types of products.
Batteries: Dispose of any end-of-life batteries according to local regulations.
Packaging waste: Ensure that any packaging waste is collected separately for
available national packaging collection and recycling services. Packaging is 80%
recyclable/renewable by weight.
5.7 Environmental Consideration of Waste
Disposal
The NXT system is designed to perform urodynamics studies. Laborie recommends
disposing of waste generated from these studies, such as urine, properly to prevent
environmental pollution. The waste should be disposed of in such a way that it will not
pollute the fresh water supply system, especially the drinking water system. In areas that
have sewage systems with water treatment procedures, the most convenient means of
disposal is to use these sewage systems.
5.8 Backups
Synergy software automatically backs up data to its database at regular intervals. You can
also archive patient data to an external hard drive. For information on archiving patient
data, refer to the section Database Management on page 55
.
5.9 Troubleshooting Guide
Symptoms Possible Cause(s) Corrective Action(s)
All Devices
No response from
the devices?
No power on electrical
outlet?
Plug the system into a known working
electrical outlet.
Damaged power cord? Unplug the system and contact Laborie
for a replacement power cord.
Power cord not
connected properly?
Ensure that the power cord is secure at
the base of system and at the electrical
outlet.