Maintenance and Service
Laborie NXT Pro Owner’s Manual LBL-00146, MAN2010 86
5.1.4.2 Disinfecting the Urocap NXT
To disinfect the Urocap NXT, follow the instructions provided in the section Performing
Manual Cleaning on page 85, allowing the surface to remain wet for the required contact
time listed on the product label of the cleaning solution.
Disinfect the area around the battery cover ventilation holes on the Urocap NXT by
performing the following steps:
Remove the vent plug if connected.
Gently wipe the affected area with a cleaning wipe to remove soiling and debris.
Take care to prevent any excess cleaning reagent from entering the vent hole.
Allow the surfaces to remain wet for the required contact time, as listed on the
product label of the cleaning solution.
Repeat this process, if necessary, using either a fresh wipe or the clean side of the
existing wipe until the affected area is visibly clean.
Reinsert the vent plug.
5.1.4.3 Disinfecting the Roam NXT and PIM NXT
To disinfect the Roam NXT, PIM NXT for T-DOC, and PIM NXT for Fluid, follow the
instructions provided in the section Performing Manual Cleaning on page 85
, allowing the
surface to remain wet for the required contact time, as listed on the product label of the
cleaning solution. For any parts that may contact the patient’s skin, such as belt clips,
complete a rinse step to aid in the removal of reagent residuals. Use a clean, lint-free cloth
that has been moistened with purified water to wipe the part. Repeat for a total of three
rinses.
Disinfect the area of the PIM NXT for T-DOC normally covered by the transducer doors by
performing the following steps:
Remove the transducer doors and safely place them aside.
Gently wipe the affected area with a cleaning wipe, being careful to avoid the
transducer channels. Take care to prevent any excess cleaning reagent from
entering the transducer channels. Wipe each transducer door to remove soiling
and debris. Allow the surfaces to remain wet for the required contact time listed
on the product label of the cleaning solution.
Repeat this process, if necessary, using either a fresh wipe or the clean side of the
existing wipe until the affected area is visibly clean.
Reattach and close the transducer doors.
5.1.4.4 Disinfecting the UPP Puller NXT
Prior to beginning the disinfection procedure, ensure that the UPP Puller NXT nose is
separated from the motor. Remove the catheter guide and the catheter clamp and disinfect
those separately. Refer to the section Assembling the UPP Puller NXT on page 24
.
To disinfect the UPP Puller NXT nose, catheter guide, and catheter clamp, follow the
instructions provided in the section Performing Manual Cleaning on page 75
, allowing the
surface to remain wet for three minutes or for the required contact time, as listed on the
product label of the cleaning solution.