Laborie NXT Pro Owner’s Manual LBL-00146 [C], MAN2010 66
4.5.1.2 Consumable Traceability
During the Preview phase, you will need to specify in the Consumable Traceability
window each of the consumables that will be used for the study.
You can do this by either scanning the SmartSense tag on the product using the
registration systems embedded in the cart’s docking station and the PIM NXT, or
by manually typing the required information. Any information that you type for
the consumables will be displayed in the Consumable Traceability panel
(Figure 54).
During scanning of the consumable, the Roam NXT (connected to the PIM NXT)
or the NXT docking station will provide auditory feedback, emitting a single beep
to indicate when a scan is in process, and an additional single beep for a
successful scan. If the scan fails, the docking station will emit two additional
beeps.
IMPORTANT: Auditory feedback is not intended to act as a standalone indicator or as
an alert for consumable scanning status. You must verify that consumables have
successfully scanned into consumable traceability by verifying their addition in the
Consumable Traceability window.
Figure 54: Consumable TraceabilityAir-Charged and Fluid Catheters
Consumables are classified as automatic and manual. Perform the steps here for scanning
the following:
Automatic consumables (e.g., the Pves catheter, Pabd catheter, or Vinf catheter
enabled with a PIM connector) are automatically recorded when the catheters are
inserted into the PIM NXT. Verify that the catheter appears in the Consumable
Traceability window.
IMPORTANT: Automatic consumables cannot be manually entered into consumable
traceability; they must be scanned by the PIM NXT.
Manual consumables (e.g., pump tubing, extension tubing, pressure measurement
tubing, Fluid NXT Catheters, or other supplies required for the study) must be
either scanned using the registration system embedded in the NXT docking