Maintenance and Service
Laborie NXT Pro Owner’s Manual LBL-00146, MAN2010 85
5.1.3.6 Performing Manual Cleaning
This section covers the cleaning method for the cart, shelves, docking station, Remote
Control NXT, and peripherals.
For the Roam NXT, PIM NXT, Urocap NXT, NXT Infusion Pump, and UPP Puller NXT, refer to
the device-specific cleaning sections starting on page 82
for guidance on when and where to
apply the cleaning procedure described here:
Before beginning to clean, fully review the sections General Cleaning and
Disinfection, Cleaning and Disinfection Preparation, Cleaning, and Completing
Cleaning and Disinfection.
Turn off and disconnect all power to the NXT system or NXT module.
Remove any visible soiling by hand immediately after the completion of a study,
using a dry, soft, lint-free cloth.
Gently wipe the surface of the cart, shelves, computer, NXT module or peripherals
using a fresh cleaning wipe to remove visible soil.
Use a fresh wipe to thoroughly wet the external surfaces of the device.
Wipe around cables, edges, cracks, crevices, corners, and hard-to-reach areas.
Ensure that all surface areas and connectors have been in contact with the wipe.
Repeat as necessary until surfaces are visibly clean. If visible soil persists, use a
soft bristle brush to gently brush away or loosen the remaining soil, and then wipe
the device again using a fresh wipe.
Wipe the cart, shelves, computer, modules, and peripherals with a dry, lint-free
cloth. Ensure that the device is dry to the touch before using it in a procedure.
Visually inspect the units in standard office lighting to ensure that they are clean.
If the inspection determines the devices to not be visually clean, the user should
repeat the cleaning cycle.
Perform visual inspection of all units for corrosion, discoloration, pitting, or
cracking. Contact Laborie for service and/or disposal instructions if the unit is
observed with these defects.
NOTE: There are no rinsing steps required for the cleaning or disinfection processes.
Disinfection
5.1.4.1 Performing Manual Disinfection
To disinfect the device, follow the cleaning instructions, allowing the surface to
remain wet for the required contact time as listed on the product label of the
cleaning solution. Refer to the module-specific cleaning instructions.
Ensure that the device is dry to the touch before using it to perform another
study.
After disinfection is complete, confirm that each module is working properly by
verifying calibration. Refer to the section
Verifying Calibration on page 90 for
calibration instructions per module.
IMPORTANT: Refer to Urocap NXT–specific disinfection instructions in the section
Disinfecting the Urocap NXT on page 86
. Refer to the disinfection instructions for Roam NXT
and PIM NXT in the section Disinfecting the Roam NXT and PIM NXT on page 86. For UPP
Puller NXT, refer to the section Disinfecting the UPP Puller NXT on page 86.