Maintenance and Service
Laborie NXT Pro Owner’s Manual LBL-00146, MAN2010 82
5.1.2.2 Urocap NXT Preparation
IMPORTANT: Ensure that the plug is fully inserted to avoid fluid ingress. Refer to
Figure 95.
5.1.2.3 Roam NXT and PIM NXT Preparation
To prepare for cleaning or disinfection, remove the belt clips from the Roam NXT and the
PIM NXT. For instructions on removing the clip, refer to the section
Removing and Attaching
the Clip on page 108.
When performing cleaning or disinfection, the Roam NXT and PIM NXT can be separated by
unfastening the screw that joins together the Roam NXT and the PIM NXT. Disconnection of
the Roam NXT and PIM NXT is not required for cleaning or disinfection. For more
information on separating the Roam NXT and the PIM NXT, refer to the section
Removing
the PIM NXT on page 107.
Identify the PIM NXT model in use. Be sure to close the transducer doors on the PIM NXT for
T-DOC.
IMPORTANT: Do not attempt to clean the inside of the PIM NXT for T-DOC transducer
channels as this may damage the delicate measurement elements. This area should not
require cleaning, given that it is protected by the transducer doors while not in use or by
the catheter connector during urodynamic study. If dirt or debris is observed in the
transducer channels as a part of daily inspection, contact Laborie for service and/or disposal
instructions.
If contamination is observed near the transducer channels during or post procedure, follow
the instructions provided in the section Cleaning the Roam NXT and PIM NXT on page 83
to
remove contamination from this area of the device.
Cleaning
Follow the instructions provided in this section to manually clean the NXT system and
peripherals.
IMPORTANT: No method of automated cleaning or disinfection has been validated for
the NXT system or its modules. None of the NXT modules, cart, shelves, computer, or
peripherals are sterilizable or autoclavable.
5.1.3.1 Cleaning the PC
To clean the PC:
1. Prepare a cleaning reagent. See the section
Cleaning and Disinfection Reagents on
page 80 for a list of reagents approved for use with the PC, per the
manufacturer’s instruction.
2. Prepare a clean cloth moistened with the cleaning solution.
3. Wipe the PC thoroughly with a clean wipe.
CAUTION: Do not immerse or rinse the PC or its peripherals. Do not spray cleaning
agents on the chassis. Do not use disinfectants containing phenol.