15.11 EasyDrive
It is possible to equip the bed with the EasyDrive wheel. The EasyDrive helps hospital sta󰀨 to drive the bed during patient transport
with minimal manpower. The EasyDrive wheel is located in the center of the bed under the undercarriage. EasyDrive is equipped
with its own accumulator and charger and it is not dependent on the bed functions so, if discharged you can still use the bed func-
tions. The bed is equipped with one EasyDrive Control Panel. EasyDrive wheel is oriented in straight direction of the bed.
15.11.1 Safety instruction for EasyDrive System
Follow the instructions carefully.
Ensure that the bed is operated exclusively by qualied sta󰀨.
Make sure the siderails are raised up during the transport.
Never use buttons for bed positioning during transport.
Never use FORWARD button when descending.
Special precaution need to be considered when reversing. Always keep distance from the bed and never use reverse
button when descending or ascending.
Do not use EasyDrive to transport on a slope over 1 degree unless adequate personnel are available to manage safe bed
transport.
Never leave the bed with an activated EasyDrive system without supervision of the trained sta󰀨.
Always use the regular mechanical brake system to brake and stabilize the bed.
Pay increased attention when driving the bed using EasyDrive. Be aware of people and objects in close proximity and
avoid collision with them by careful driving, especially by appropriate speed control.
Make sure the bed is unplugged and bed brakes are released before using EasyDrive.
Retract the EasyDrive wheel to the undercarriage by adjusting the neutral pedal position. This will prevent misuse when
unbraking and braking the bed.
Switch o󰀨 the EasyDrive and ask the authorised service technician for disconnecting the EasyDrive accumulator prior to
long-term storage or transport.
Press ON/OFF button on the EasyDrive control panel in the case of EasyDrive system failure.
Retract the EasyDrive wheel to the undercarriage every time you intend to move the bed sideways.
Pay attention to the Accumulator Charge Status Indicator and plan your drive using the EasyDrive accordingly. Insu󰀩cient
accumulator capacity can cause unexpected complications and risks during the drive.
Always plug the bed in when you nish your drive in order to recharge the accumulator and keep your bed ready to go
using the EasyDrive.
The EasyDrive accumulator must be replaced every 2 years to maintain proper functions of the EasyDrive.
15.11.2 Intended Use of the EasyDrive System
15.11.3 Compatible Castors
EasyDrive system is intended to facilitate the transport of a compatible bed with patient by the medical sta󰀨 using the fth 
driven wheel of the EasyDrive system. EasyDrive system is not intended to stop the bed in an emergency case.
Unintended use:
riding the bed
other usage than described in instructions for use
by other person than the trained sta󰀨
EasyDrive system is compatible with Tente Integral Soft brake simple castors with diameter 150 mm.
Essenza 300 medical bed can transport only single patient at a time and cannot be used to transport other items (except bed acce-
ssories in secured position).