D9U001EZ3-0101_00174
17.5 Infusion Pump Holder
17.6 Infusion Bottle Basket
Fig. Infusion Bottle Basket Fig. Infusion Pump Holder at bed head end
Infusion Bottle Basket is intended for carrying bottles with intrave-
nous solutions.
It is intended to be suspended on a compatible Infusion Stand.
Maximum Load of the Infusion Bottle Basket is 2 kg.
Infusion Pump Holder is intended for holding a compatible infusion pump.
Infusion Pump Holder must be placed into a bushing at head end of the mattress support platform or into a bushing on the Acces-
sory Adaptor at bed head end.
WARNING!
Risk of injury due to incorrect use!
► Ensure the Infusion Pump Holder will not collide with any movable parts of the bed (especially Backrest bed
part) or with the patient. This must be veried during installation.
► Do not overtighten the infusion pump clamps during tment. Overtightening may damage the Infusion Pump
Holder.
► Infusion pump can only be used if the Infusion Pump Holder is tted into a bushing at head end of the mattre-
ss support platform or into a bushing on the Accessory Adaptor at bed head end.
► Do not use the Infusion Pump Holder as driving/pushing device during the bed transport.