5.2.2 Mechanical Specications of the Essenza 300 with Lateral Tilt
Parameter Value
External Bed Length 219 cm (221 cm for bed with Accessory Adaptor with 2
Bushings for Lifting Pole at bed head end)
External Bed Width 99,5 cm
Maximum Siderail Height above Mattress Support Platform:
with plastic siderails
with telescopic siderails
with collapsible siderails
42 cm
43 cm
38 cm
Siderail Dimensions (length x height):
plastic siderails (head siderail; foot siderail)
telescopic siderails (head siderail; foot siderail)
collapsible siderails
101,3 cm; 88,5 cm x 43,5 cm
100 cm; 100 cm x 43 cm
162 cm x 38 cm
Maximum Mattress Dimensions (length x width) 200 cm x 90 cm
Maximum Mattress Height:
with plastic siderails
with telescopic siderails
with collapsible siderails
20 cm
20 cm
16 cm
Bed Extension 0 cm / 20 cm / 30 cm
Calfrest Extension 0 cm – 20 cm
Castor Diameter:
125 Tente Integral simple castor
125 Tente Linea double castor
150 Tente Linea double castor
150 Tente Integral Soft brake simple castor
12,5 cm
12,5 cm
15 cm
15 cm
Minimum — Middle — Maximum Mattress Support Platform
Height Adjustment above oor (without mattress):
Essenza 300 with Lateral Tilt with 125 Tente Linea castors
Essenza 300 with Lateral Tilt with 150 Tente Linea castors
Essenza 300 with Lateral Tilt with 150 Tente Integral Soft
brake simple castors
39,4 cm – 51,3 cm – 89,8 cm
41,9 cm – 53,8 cm – 92,3 cm
44,7 cm – 56,6 cm – 95,1 cm
Maximum Clearance under the Undercarriage in Standard
Position (with 125 Tente Linea castors)
12,4 cm
Maximum Backrest Angle 70°
Maximum Thighrest Angle 35°
Maximum Calfrest Angle 25°
Maximum Lateral Tilt Angle 15°
Maximum Angle between Calfrest and Thighrest 300°
Ergoframe Distances (Backrest Distance / Thighrest Distan-
ce)
5 cm / 3 cm
Maximum Trendelenburg Tilt Angle 14°
Maximum Reverse-Trendelenburg Tilt Angle 14°
Average Bed Weight 180 kg
Maximum Lifting Pole Load 75 kg
Safe Working Load (SWL) 270 kg
Maximum Mass of Mobile Hospital Bed (Maximum Mass of
Empty Bed + Safe Working Load)
470 kg
Maximum Patient Weight
Application environment 1, 2
Application environment 3, 5
205 kg
235 kg
Maximum Sound Pressure Level 53 dBA