Parameter Value
Input Voltage, Frequency:
version without integrated mattress
version 1 with integrated mattress
version 2 with integrated mattress
version 3 with integrated mattress
version 4 with integrated mattress
version 5 with integrated mattress
100 — 240 V AC, 50/60 Hz
100 — 127 V AC, 60 Hz
220 — 240 V AC, 50 Hz
220 — 240 V AC, 60 Hz
100 V AC, 50 Hz
220 V AC, 60 Hz
Maximum Power Input max. 480 VA
Ingress Protection according to EN 60529 IPX4
Electric Protection Class Class I (with type B Applied Parts)
Duty Cycle max. 2 minutes ON / 18 minutes OFF
Bed Accumulator AKU Pb VRLA 2 x 12 V / 1,2 Ah / Fuse 15A
EasyDrive Accumulator AKU Pb 24 V / 7 Ah
Control Unit Fuses:
version without integrated mattress
version with integrated mattress
2x T4A L 250 V
2x T4A L 250 V, 2x T1A L 250 V
5.3 Electrical Specications
5.4 Environment Conditions
CAUTION!
Risk of damaging the product if its packaging is exposed to environmental conditions outside of those speci-
fied in the Environment Conditions chapter!
WARNING!
Risk of damaging the product due to incorrect environment conditions!
Do not use the Essenza 300 bed under the environmental conditions outside of those specified in the
Environment Conditions chapter!
Parameter Value
Use Conditions
Ambient Temperature 10°C — 40°C
Relative Humidity 30% — 75 %
Atmospheric Pressure 795 — 1060 hPa
Storage and Transport Conditions
Ambient Temperature -20°C — 50°C
Relative Humidity 20% — 90 %
Atmospheric Pressure 795 — 1060 hPa