D9U001EZ3-0101_0052
The bed is equipped with a standard protective connector. This connector is used for potential equalisation between the bed and
any intravascular or intracardiac device connected to the patient to protect the patient from static electric shocks.
Use equalisation connector if the patient is connected to any intravascular or intracardiac device.
Before connecting the patient to an intravascular/intracardiac device:
Connect the ground wire of the device to the potential equalisation connector on the bed on which the patient in question
is lying.
Use a standard hospital connector.
Make sure that the connectors match.
Make sure that there is no possibility for inadvertent disconnection.
8.8 Before Use
Prepare the bed for use as follows:
Connect the bed to the mains.
Charge the accumulator.
Raise and tilt the mattress support platform to the highest position.
Lower and tilt the mattress support platform to the lowest position.
Check that the castors as well as brake levers work correctly.
Check that the bed extension works correctly.
Check that it is possible to remove the head board and foot board.
Check all of the functions on the control elements.
Check that the siderails function properly.
Dispose of all packaging (see „21 Disposal“ on the page 199).
8.9 Transport
For a safe transport, observe the following:
Ensure that no cables are run over when moving the bed.
Ensure that the power supply cord is attached with a hook (at the head end of the bed).
Adjust bed height to at least 20 cm below maximum height.
Push bed by handles on Head Board or Foot Board.
Move the bed exclusively on suitable oor surfaces.
Ensure the bed is braked when it should not move (see „11.4 Castor Control“ on the page 68).
For longer distances, ensure the castor direction determining function is activated.
Ensure that the castors are not braked before moving the bed.
Suitable surfaces:
Tile
Hard linoleum
Poured ooring
Unsuitable surfaces:
Too soft, unsealed or defective ooring
Soft wooden ooring
Soft and porous stone oors
Carpeted oors with underlay
Soft linoleum
Before moving the bed:
Disconnect the patient from the intravascular or intracardiac device.
Disconnect the potential equalisation connector.