EasyManua.ls Logo

Micromeritics TRISTAR II PLUS - Page 145

Micromeritics TRISTAR II PLUS
310 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IMPORT ASCII TEXT DATA
ASCIItext file format rules
n
The header must consist of one line to include title, two unit specifications, and distribution
type:
o
Accepted pore dimension units are: A, nm, um
o
Accepted pore volume units are: cm
3
/g, cm
3
/g, mL/g
o
Accepted distribution types are: cumulative, incremental
Two examples of a header format:
My Title (A, cm
3
/g, incremental)
My Title (A, cm
3
/g, cumulative)
n
The data must be in two columns and should be separated by a comma or white space.
n
The data lines must be ordered so that pore dimensions are monotonically increasing or
decreasing.
Window appearance will vary depending on the selected report. This function can be performed
on samples files with a Complete status or during an analysis.
1. Create the ASCIItext file using the proper format as indicated above.
2. In the analyzer application, go to File > Open.
3. Select a sample file to overlay graphs on to.
4. Click Open (or double-click the file name).
5. Right-click in the graph area and select Edit imported data.
Generate Pore Size Distribution Graph Overlays
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) Sep 2023
7 - 29