Selections
Description
Pressures [button]
Use to select a pressure range for report calculations and points for
exclusion from calculations.
Calculation pressure range. Enter the minimum and maximum
pressures to be used in the pressure table.To exclude a point from
the calculations used to generate the report, select Exclude.
Exclude All.Select to exclude all pressure points in the table.
Include All.Select to include all pressure points in the table.
Regularization
[drop-downbox]
Select the extent of smoothing to apply to the data. If 0.20000 (user)
is selected, enter a number in the text box giving a relative weight for
the smoothing during deconvolution. Larger values produce more
smoothing.
Select Isotherm Data
[groupbox]
Select the isotherm data to be used in the report.
Select Reports
[groupbox]
Select the reports to generate. To edit graph details, highlight the
graph option and click Edit. The Log Goodness of Fit and Goodness
of Fit graphs cannot be edited.
Autoscale Options. Use to autoscale the x-axis and/or y-axes.
DFTPore Size Report (continued)
DFT Pore Size Report
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) — Sep 2023
8 - 15