Revision:1.0(2023-01-12)
223
1 One flathead screw
2 Cover of the module
3 SSD card
7.15.3 Commissioning and Verification
Precautions
None
Tools Required
USB flash drive
Fixtures Required
None
Precondition:
Before replacing the SSD card, you need to provide the serial number and MAC address of the
machine, obtain the Key file of the machine from Mindray, and decompress the Key file to the
USB flash drive for backup.
Specific Steps
1. After replacing SSD or HDD, contact Mindray after-sales service engineer to download the
configuration file from Mindray internal platform (CRM or Ukmo).
2. Decompress the configuration file, copy it to a USB drive and check whether the folder contains
Config.key.