76
3. Unpack the device in a spacious place and check the items in the package against the packing
list.
The device must be inspected in the presence of physicians from the ultrasound
department and personnel from the equipment department or other personnel
authorized by the hospital.
The person in charge of product acceptance needs to sign on the packing list after
verifying that the items in the package are consistent with those on the packing list.
The product will be considered non-conforming in any of the following cases:
The package is damaged during transportation.
Wrong labels are found on the device.
Inconsistency between the packing list and the product configurations specified
in the contract
Inconsistency between the packing list and actual articles, with accessories
being absent
Incomplete equipment parameters or functions, resulting in an installation failure
due to faults and thereby an acceptance failure
Equipment problems resolved by installation personnel to ensure successful
installation
Any of the preceding problems is found by the distributor before the device is
delivered to the customer.
3.2.3 Unpacking Tools
Name Type Part Number Remarks
Scissors or diagonal
pliers
/ / /
3.2.4 Unpacking the Device
1. Cut the straps with scissors or diagonal pliers.
Diagnostic Ultrasound System
Service Manual
3 Installation