Revision:1.0(2023-01-12)
276
1. Refer to the disassembly procedure in 7.47.2 Disassembly and Assembly to remove the battery
box cover. Remove the fixing screws from the battery box assembly and take off the battery box
assembly.
2. Refer to the disassembly procedure in 7.38.2
Disassembly and Assembly to remove the battery
box bracket and bottom cover.
3. Remove the two M3x8 Phillips pan-head screws and unplug the cable terminals, to take off the
auxiliary output holder of the trolley (including the auxiliary output socket).
1 Two M3x8 Phillips pan-head screws
2 Cable terminal
3 Trolley Auxiliary Output Holder (Including Auxiliary Output Socket)
7.46.3 Commissioning and Verification
None
7.47 Battery Box cover (FRU)
7.47.1 General Information
Diagnostic Ultrasound System
Service Manual