224
3. Insert the USB drive to the USB interface, and start the machine. The system prompts
importing the configuration file.
4. Tap Import Configuration, select the key folder in the USB drive, and then tap OK. After the
file is imported, the system prompts "Imported successfully".
5. After the machine is started, enter the About screen.
6. Check ECN/TCN on the About page. If Incorrect is displayed, the board logic or firmware
does not match. Therefore, the Doppler software needs to be restored.