EasyManua.ls Logo

Mindray TEX20

Mindray TEX20
445 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Revision:1.0(2023-01-12)
224
3. Insert the USB drive to the USB interface, and start the machine. The system prompts
importing the configuration file.
4. Tap Import Configuration, select the key folder in the USB drive, and then tap OK. After the
file is imported, the system prompts "Imported successfully".
5. After the machine is started, enter the About screen.
6. Check ECN/TCN on the About page. If Incorrect is displayed, the board logic or firmware
does not match. Therefore, the Doppler software needs to be restored.
7.16 Wi-Fi module Wi-Fi + BT
7.16.1 General Information
Diagnostic Ultrasound System
Service Manual
7 FRU Replacement

Table of Contents

Other manuals for Mindray TEX20

Related product manuals