7.22 Hook cover (sy)
7.22.1 General Information
FRU Code FRU Name Lower-Level FRU
Code
Lower-Level FRU
Name
043-014654-00 Hook cover (sy)
7.22.2 Disassembly and Assembly
Precautions
Protect the exterior surface of the parts.
Tools Required
Art knife, crosshead screwdriver (M4), and flathead screwdriver
Fixtures Required
None
Specific Steps
1. Use an art knife to remove the four screw hole identification stickers on the rear of the hook
decoration cover.
2. Use a crosshead screwdriver to remove the four M4X12 crosshead combination screws that
connect the hook decoration cover.
3. Use a flathead screwdriver to push up along the lower edge of the hook decoration cover to
remove the hook decoration cover.