Revision:1.0(2023-01-12)
70
3 Installation
3.1 Site Preparations
3.1.1 Environmental Requirements
33..11..11..11 Environmental Requirements
Operating Environment Storage and transportation
environment
Ambient temperature 0°C to 40°C –20°C to +55°C
Relative humidity 20%-85% (no condensation) 20%-95% (no condensation)
Atmospheric pressure 700 hPa to 1060 hPa 700 hPa to 1060 hPa
Do not use, store, or transport the system beyond the specified environmental
conditions.
Do not place the system in any of the following environments:
• Location near a heat source
• High-humidity environment
• Environment with flammable gases
3.1.2 Power Supply Requirements
33..11..22..11 Power Supply Requirements
Power specifications
Power input
Voltage 100–240VAC
Frequency 50/60Hz
Power consumption 150 VA
Power output
Diagnostic Ultrasound System
Service Manual
3 Installation