Revision:1.0(2023-01-12)
229
FRU Code FRU Name Lower-Level FRU
Code
Lower-Level FRU
Name
115-083590-00 Dustproof Filter (FRU)
Function description:
It prevents dust entering the main unit.
7.19.2 Disassembly and Assembly
Precautions
Protect the exterior surface of the parts.
Tools Required
None
Fixtures Required
None
Specific Steps
1. Refer to the disassembly procedures in 7.2.2 Disassembly and Assembly, 7.39.2 Disassembly
and Assembly
, and 7.40.2 Disassembly and Assembly to take off the display assembly.
2. Remove the main unit from the trolley, put it on the table, and take off the dust-proof net from
the cover of the dust-proof net.