Revision:1.0(2023-01-12)
174
Button or Check Box Description
Save Report Click this button to save the test data to the root directory under the
specified logical drive letter. This button is unavailable during the test
or if no logical drive letter is selected.
Shut Down Click this button to close the program, and the machine shuts down
automatically. If the test report is not saved, the system will prompt you
whether to save the test report to the mobile storage medium before
shutdown.
When you enter the user self-test page, the self-test starts directly. After the test is complete, if
any test item fails, the page only displays the ID of the failed test item. For example, when
"Parameter board I2C communication test" fails, only Z0801 is displayed on the screen.
According to the appendix, the test item corresponding to Z0801 is "Parameter board I2C
communication test".
The PASS dialog box pops up only when the test results of all test items are PASS, as shown in
the figure below.
55..22..55..44 Test Report
Test report
The test report is in HTML format and can be opened using Microsoft Internet Explorer. The
format of the maintenance self-test report is as follows:
Diagnostic Ultrasound System
Service Manual
5 Maintenance