150
5 Maintenance
5.1 Care and Maintenance
5.1.1 Overview
55..11..11..11 Description
The ultrasound system does not require regular preventative maintenance such as calibration or
adjustments to the hardware and transducers. Periodic cleaning and quality inspection described
in this chapter are recommended.
55..11..11..22 Tools, Measurement Devices and Consumables
Table 5–1 Tools and measuring devices
Tool or Measuring Device Quantity Remarks
Plastic or resin container 1 Used to store the saline, and
can accommodate two
ultrasound probes.
Soft brush 1 Its size is similar to that of a
toothbrush, and it has a soft
head.
Small plastic basin 1 Used to store the soapy
water.
Safety tester 1 Used when conducting
electrical safety tests. For
details, see 9.4
Electrical
Safety Maintenance
Table 5–2 Tools and measuring devices
Consumables Quantity
Remarks
Aluminum foil About 1 m /
Normal saline solution About 1000 mL At least half of the container
should be filled with the
solution so that the probe can
be submerged by the solution.
(Concentration: 0.85-0.95%)
Diagnostic Ultrasound System
Service Manual
5 Maintenance

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals