Revision:1.0(2023-01-12)
329
Fault Description
User Permission of Viewing Errors Fault Description Corresponding to the
Permission
User and Administrator Permission None
Service Engineer Permission None
Log Record
"Date" "Time" "Errocode": System Monitor:Power supply alert! [XXX], Current voltage: [VVV] V,
Limit voltage: [LLL]-[HHH]V.
Involved FRU
Main board assembly
Severity
Maintenance required
Handling Level
Contact the device provider for onsite handling.
Error Mechanism
The system detects that the corresponding power voltage on the main board does not meet the
requirements.
Possible Cause
The corresponding power circuit on the main board fails.
Resolve Path
If the errors related to this voltage is consistently reported in logs, replace the main board.
Solution
Replace the main board. For details, refer to 7.13.2 Disassembly and Assembly.
88..22..22..1144 V033: System Monitor: Power supply alert!
Failure Code
V033
Diagnostic Ultrasound System
Service Manual
8 Troubleshooting

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals