331
T1001
Fault Description
User Permission of Viewing Errors Fault Description Corresponding to the
Permission
User and Administrator Permission Temperature alarm. Please clean the dust-
proof net.
Service Engineer Permission
Temperature alarm. Please clean the dust-
proof net.
Log Record
"Date" "Time" "Errocode": System Monitor: Temperature Alert!please clean the dust-proof cover,
Current Fan Speed: [FFF]rpm
Involved FRU
/
Severity
Maintenance as soon as possible
Handling Level
Handle the error yourself.
Error Mechanism
The system detects that the FPGA temperature on the main board is close to the warning value.
Possible Cause
1. The ambient temperature for the device is too high.
2. There are sundries around the air inlet and outlet of the device, causing poor ventilation and
dissipation.
3. The dust-proof filter inside the device is blocked, which prevents heat from being discharged.
Resolve Path
Confirm the ambient temperature and ventilation and dissipation of the device and clean the dust-
proof net.
Solution