"Date" "Time" "Errocode": System Monitor: Temperature Alert!please clean the dust-proof cover,
Current Fan Speed: [FFF]rpm
Involved FRU
/
Severity
Maintenance as soon as possible
Handling Level
Handle the error yourself.
Error Mechanism
The system detects that the SSD temperature is close to the warning value.
Possible Cause
1. The ambient temperature for the device is too high.
2. There are sundries around the air inlet and outlet of the device, causing poor ventilation and
dissipation.
3. The dust-proof filter inside the device is blocked, which prevents heat from being discharged.
Resolve Path
Confirm the ambient temperature and ventilation and dissipation of the device and clean the dust-
proof net.
Solution
1. If the ambient temperature is too high, move the machine to an area with a lower ambient
temperature (such as an air-conditioned room).
2. Remove the sundries from the air inlet and outlet of the device to ensure good ventilation and
dissipation.
3. Remove the dust-proof filter inside the device and perform cleaning and maintenance.
88..22..33..55 T1000: FPGA temperature alarm. Please save the current data
quickly. Shut down in xx seconds.
Failure Code
T1000
Diagnostic Ultrasound System
Service Manual
8 Troubleshooting

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals