Revision:1.0(2023-01-12)
337
2. Check whether the FPGA heat dissipation module on the main board is firmly connected to
FPGA, and the heat-conducting cream is spread evenly.
3. Check whether the fan works properly.
4. If the above factors are eliminated, replace the main board.
Solution
1. If the ambient temperature is too high, move the machine to an area with a lower ambient
temperature (such as an air-conditioned room).
2. Remove the sundries from the air inlet and outlet of the device to ensure good ventilation and
dissipation.
3. Remove the dust-proof filter inside the device and perform cleaning and maintenance.
4. Check whether the fan works properly. If the fan works at a normal speed but the fault persists,
replace the main board. For details, refer to 7.13.2
Disassembly and Assembly.
88..22..33..66 T006: CPU temperature alarm. Please save the current data quickly.
Shut down in xx seconds.
Failure Code
T006
Fault Description
User Permission of Viewing Errors Fault Description Corresponding to the
Permission
User and Administrator Permission
CPU temperature alarm. Please save the
current data quickly. Shut down in xx seconds.
Service Engineer Permission CPU temperature alarm. Please save the
current data quickly. Shut down in xx seconds.
Log Record
"Date" "Time" "Errocode": System Monitor:Temperature alert! [XXX], Current temperature: [VVV]
C, Current Fan Speed:[FFF]rpm
Involved FRU
PC module assembly
Severity
Suspension
Diagnostic Ultrasound System
Service Manual
8 Troubleshooting