Revision:1.0(2023-01-12)
344
Solution
1. If the ambient temperature is too high, move the machine to an area with a lower ambient
temperature (such as an air-conditioned room).
2. Remove the sundries from the air inlet and outlet of the device to ensure good ventilation and
dissipation.
3. Remove the dust-proof filter inside the device and perform cleaning and maintenance.
4. If the fault persists after the preceding measures are taken, replace the SSD module. For
details, see 7.15.2
Disassembly and Assembly.
8.2.4 Fan Alarms
88..22..44..11 F1001: System fan requires maintenance. Please contact the
service engineer in time.
Failure Code
F1001
Fault Description
User Permission of Viewing Errors Fault Description Corresponding to the
Permission
User and Administrator Permission System fan requires maintenance. Please
contact the service engineer in time.
Service Engineer Permission System fan requires maintenance. Please
contact the service engineer in time.
Log Record
Date "Time" "Errocode": System Monitor: Fan alert! [XXX], Current speed: [VVV] rpm, Lower Limit
speed: [LLL] rpm.
Involved FRU
Main Unit Dual Fans assembly
Severity
Maintenance as soon as possible
Handling Level
Handle the error yourself.
Diagnostic Ultrasound System
Service Manual
8 Troubleshooting

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals