Severity
Maintenance as soon as possible
Handling Level
Handle the error yourself.
Error Mechanism
The system detects that the speed the main unit fan FAN1 is lower than the standard value.
Possible Cause
1. The fan is blocked due to sundries.
2. The fan is faulty.
Resolve Path
Check whether there are sundries around the fan and clean and maintain the fan. If the error
persists, replace the main unit fan assembly.
Solution
1. Remove the main unit fan assembly and clean and maintain the fan.
2. If the fault persists after cleaning, replace the Main Unit Dual Fans assembly. For details, refer
to 7.17.2
Disassembly and Assembly.
88..22..44..33 F1003: System fan requires maintenance. Please contact the
service engineer in time.
Failure Code
F1003
Fault Description
User Permission of Viewing Errors Fault Description Corresponding to the
Permission
User and Administrator Permission System fan requires maintenance. Please
contact the service engineer in time.
Service Engineer Permission System fan requires maintenance. Please
contact the service engineer in time.