363
Possible Cause
1. The PHV voltage load causes the abnormal voltage output.
2. The power circuit of the PHV board fails.
Resolve Path
Cross-test the probe and the probe groove. If the probe and the probe board are normal, apply to
replace the main unit assembly.
Solution
1. Restart the machine, and check whether the fault is rectified.
2. Check whether the fault exists on a specific probe groove. If yes, clean the probe groove. If the
fault persists after cleaning, replace the probe board. For details, refer to 7.9.2
Disassembly and
Assembly
.
3. Check whether the fault exists on a specific probe. If yes, replace the probe.
4. If the fault appears on all probes, cross-test or replace the main board. For details, refer to
7.13.2
Disassembly and Assembly.
88..22..55..1111 P017: Transmit voltage error. The image cannot be displayed
correctly. Please try to reboot the device.
Failure Code
P017
Fault Description
User Permission of Viewing Errors Fault Description Corresponding to the
Permission
User and Administrator Permission Transmit voltage error. The image cannot be
displayed correctly. Please try to reboot the
device!
Service Engineer Permission Transmit voltage error. The image cannot be
displayed correctly. Please try to reboot the
device!
Log Record
"Date" "Time" "Errocode": System Monitor:HV supply alert! [XXX] Voltage protect.
Involved FRU
Diagnostic Ultrasound System
Service Manual
8 Troubleshooting